Mass loss, molecular weight and erosion

Most of the error in mass loss and molecular weight readouts enters at the bench, before the instrument. GPC needs a stated calibration, because polystyrene standards give PLGA a relative molecular weight. Dry mass needs buffer salt dealt with, since standard PBS carries nearly 10 g of salts per liter. Wet mass needs the same blot or drain, at the same timing, every time.

Where does each readout’s error come from?

Each readout has its own prep step, and that step carries its main error source. In Lin’s in situ implants, some day-7 samples prepared for SEM lost their structure in the freeze-dryer. Polymer degradation testing in depots matches each readout to the question it answers.

Readout Prep step Main error source How to control it
Molecular weight (GPC) Freeze, freeze-dry, dissolve at a set concentration, filter Calibration: polystyrene standards give relative values One method for the whole run, normalized to its own baseline; light scattering with measured dn/dc for absolute values
Dry mass Rinse or correct for salt, freeze, dry to constant weight Buffer salt, residual water or solvent, crumbs lost in handling One salt treatment for every sample, repeat weighings until stable, dry and weigh in the same tared vial
Wet mass Blot or drain, weigh at a fixed time Surface water that varies with blot pressure and drain time; fragments lost at each weighing One method and one timing for every sample; catch what falls off
Erosion by volume CT or MRI, segmented by a threshold Threshold choice; swelling adds volume while polymer leaves One threshold rule for every sample; read volume beside mass

How do you measure molecular weight on a depot sample?

Pull the depot, freeze it, freeze-dry it, then dissolve it at a fixed concentration and inject it on GPC. The calibration decides whether the number is relative or absolute. Park and colleagues note that PLGA and polystyrene take different sizes in the same solvent, so a polystyrene-calibrated PLGA value isn’t accurate. They recommend PLGA standards of known absolute molecular weight, matched to the lactide to glycolide ratio.

Two published depot methods show the choice:

  • Relative: Lin and colleagues freeze-dried implants for 24 hours, dissolved them in THF at 1.5 mg/mL and calibrated against polystyrene.
  • Absolute: Hopkins and colleagues froze implants at −80 °C, freeze-dried them for four days and ran 2.5 mg/mL in acetone through light scattering, refractive index and viscometer detectors. They chose acetone for its larger dn/dc for PLGA than THF, and measured dn/dc for each lactide to glycolide ratio.

Read the distribution as well as the average. Mn counts chains equally, so a burst of short fragments pulls it down fast. Mw weights each chain by its mass. Dispersity is Mw divided by Mn. Random scission alone pushes it toward 2, fragments leaving the sample narrow it, and chains breaking faster in one region than another can split the trace. Lu, Garcia and Mikos saw heterogeneous chromatograms in thick PLGA films, a split that a single Mn would have hidden.

Normalize to your own baseline. Hopkins normalized Mn to the day-1 implant. A relative value is still useful for a trend, as long as every timepoint ran on the same method.

How do you get a clean dry mass?

Deal with the salt, dry the sample until its weight stops changing, and weigh it in the vessel it dried in. Each step closes one error.

Salt. Standard PBS carries 9.88 g of salts per liter (8 g sodium chloride, 0.2 g potassium chloride and 1.68 g of phosphates), and Lu’s films sat in 0.2 M PBS. If a depot holds 100 µL of standard PBS, our assumed volume, it dries to about 1 mg of salt by our arithmetic, which is 4% of a 25 mg dry depot. None of the Hopkins, Fisher or Lin methods describes a rinse before drying. That leaves two routes:

  • Rinse briefly in water: this can carry out drug or soluble fragments. Schliecker and colleagues found D,L-lactic acid oligomers under 800 Da dissolved in pH 7.4 buffer but not in water. By our reasoning, a brief water rinse is less likely to strip those oligomers than more buffer would be, though their study says nothing about glycolide-rich fragments.
  • Correct instead of rinsing: subtract the salt carried by each sample’s measured water uptake at the buffer’s salt content.

Either way, treat every sample the same, the baseline included.

Water and solvent. Dry to constant weight rather than to a clock, and don’t read constant weight as proof that solvent has gone. Some of Lin’s day-7 SEM samples still held enough residual solvent to collapse after 24 hours of freeze-drying. Makadia’s review, citing earlier work, reports that PLGA’s glass transition falls as molecular weight falls, so a degraded sample softens at a lower temperature than the starting polymer. By that logic, a heated vacuum oven is riskier late in a run than early.

Handling. Fisher’s freeze-dried samples from day 40 on were mostly powder that crumbled when handled. Weigh in a tared vial, and don’t transfer the sample.

The baseline matters too. In a solvent-based depot, a percentage of injected mass includes solvent, which polymer degradation testing covers with Hopkins’s 39% example.

How do you weigh wet mass reproducibly?

Choose blotting or draining, then hold that method and its timing fixed for every sample and every timepoint. Fisher and colleagues gently blotted their scaffolds dry before each weighing, five samples per group. After about day 15, individual samples diverged.

  • A blot: takes off surface water, so the reading sits closer to the depot and the water inside it. Pressure, paper and contact time vary between people and between days.
  • A drain: leaves a surface film. By our reasoning, untested in the sources here, a fixed drain time makes that film more repeatable than a hand blot, though the film grows as an eroding depot gains surface area and pores.
  • Timing: a wet sample loses water to the air from the moment it leaves the medium. Weigh at the same interval after removal every time.
  • Losses: a wet sample goes back in the bath, so anything it sheds is gone from every later reading. Catch fragments, or record when shedding starts.

The swelling ratio calculation from wet and dry mass is on the hydrogel swelling ratio page.

How is erosion measured besides mass?

Erosion is material leaving, so mass loss is its direct measure, and volume and imaging show where it’s leaving from. Lin and colleagues measured implant volume from 3D CT images, with a threshold set by visual assessment somewhere between 25,000 and 50,000 counts, under the same criteria for every sample. Volume and weight rose together, then fell. Fisher scanned freeze-dried scaffolds by micro-CT for porosity and pore size.

Volume isn’t mass. A swelling depot can gain volume while it loses polymer, so a volume curve needs a mass curve beside it. Zolnik and Burgess used morphology studies to show surface pitting and channels at pH 7.4, which a mass curve can’t show.

How many samples does a destructive readout use?

Count timepoints times replicates for each destructive readout. By our arithmetic, 8 timepoints in triplicate use 24 depots. Hopkins and colleagues took their dry mass baseline from each implant’s mass at injection, which destroys nothing, and ran GPC on the same freeze-dried implants they weighed, so one set of 24 gave both. A dried day-0 set adds 3 more, and mechanics needs its own set. A wet mass curve uses its own samples for the whole run. We’d add a spare per timepoint to cover a crumbled or dropped sample.

Check the balance against the sample, too. By our arithmetic, the readout alone sets a floor: reading mass loss to 0.1% of the sample takes at least 10 mg on a balance that reads 0.01 mg, and 100 mg on one that reads 0.1 mg. Mass loss is the difference of two weighings, each with its own error, so the real requirement is larger.

What a GelTech study measures on these readouts

GelTech weighs depots on Carousel, six samples at a time, tracking drained wet mass on the same samples at every timepoint, so the wet curve spends no destructive set. Carousel reads to 1 mg, and its minimum dry sample is 25 mg. By our arithmetic, 1 mg is 4% of that smallest sample, so a small depot’s curve shows trend and timing rather than the 0.1% floor above. GelTech runs molecular weight in its own lab on separate samples, and integrated instruments read drug in the sampled medium, so depot mass and drug release sit on one timeline. PLGA degradation and release shows what that pairing explains.

GelTech doesn’t run animal or clinical studies, or accredited biocompatibility or sterility testing. See drug delivery and what to put in a study brief.

Sources

Related reading

GelTech services and industries

Questions

Should PLGA molecular weight be reported as Mn or Mw?

PLGA molecular weight is best reported as both Mn and Mw, with dispersity, because each moves differently as chains break. Published depot work has used either one alone: a 2019 in situ implant study normalized Mn to day 1, and a 2007 microsphere study fitted Mw to first-order kinetics.

Can inherent viscosity replace GPC for PLGA?

Inherent viscosity can stand in for GPC as a single molecular weight number. Suppliers specify PLGA lots by an inherent viscosity range, measured in hexafluoroisopropanol or chloroform, and convert it to molecular weight through a standard curve for that polymer and solvent pair, a 2021 review notes. It gives no distribution, so it can't show dispersity or a split population.

How long should a degradation sample be freeze-dried?

A degradation sample should be freeze-dried until its weight stops changing, rather than for a fixed time. Published PLGA implant studies used 24 hours and four days, and a sample that still holds water or solvent at a fixed cutoff reads as less mass loss than it had.