MVTR for wound dressings
MVTR is the mass of water vapor a wound dressing lets through per unit area in 24 hours. EN 13726:2023 measures it two ways, with vapor against the dressing (upright) or liquid against it (inverted), and the two can disagree. In one 3M-funded study, an IV film read 1,682 g/m²/24 h upright and 4,089 inverted, while another film read about the same either way.
What does MVTR measure in a wound dressing?
MVTR measures how fast water vapor crosses a dressing from the wound side to the air. EN 13726:2023 defines it as the “rate of permeability of materials to the passage of moisture vapour molecules from the skin contact side to the external atmosphere under controlled conditions of humidity and temperature.” Its scope names MVTR methods for “permeable film wound and fixation dressings.”
Check the unit before you compare two numbers. The note to the 2023 definition expresses MVTR in grams per square centimeter of absorbent area per 24 hours. The MVTR annexes’ calculation clauses aren’t in the public preview, and definition 3.5 calls a film dressing non-absorbing, so the public text leaves open how “absorbent area” applies to a film. Papers that used the 2002 method report grams per square meter per 24 hours. By our arithmetic, 1 g/cm² is 10,000 g/m², so 1,682 g/m²/24 h is about 0.17 g/cm²/24 h. Why vapor loss matters to healing is covered in wound dressing fluid handling.
How does the EN 13726 cup method work?
The cup method seals a dressing over a cup of fluid, holds it at a set temperature and humidity, and weighs the cup to see how much water left through the dressing. Zehrer and colleagues, all 3M employees, ran the 2002 method on a Paddington cup with a 10 cm² opening, at 37 ± 1 °C and 19% relative humidity in a chamber with a circulating fan. Bainbridge and colleagues, also 3M-funded, describe the 2002 standard as calling for 20 mL of water and a 3.57 cm aperture.
The 2023 edition changed the hardware. Its Annex A gives the reasons:
- Doming: under the 2002 fluid handling and MVTR methods, some newer dressings passed vapor so fast that they pulled a partial vacuum and stretched, overstating fluid handling capacity. A vented lid with a 0.25 mm hole now equalizes pressure for Annexes E and I. Annex A doesn’t list a vented lid for the upright test, Annex H.
- Sealing: the test cylinder has its own annex (M), with bolts tightened to 2 Nm.
- Airflow: a minimum 2 cm gap between sample and incubator shelf.
- Fluid: Test Solution A, specified in Annex L, for both MVTR annexes. The 2002-method papers here used water.
The 2023 temperature, humidity and duration for Annexes H and I sit in the annexes themselves, past the end of the public preview.
Why do upright and inverted MVTR differ?
Upright and inverted MVTR differ because liquid against a film can change the film. In the upright test, vapor crosses an air gap before it reaches the dressing. In the inverted test, fluid sits on it. Bainbridge and colleagues had seven IV films tested both ways at an independent lab, with a smaller 2 cm aperture, 10 mL of water and 37 ± 2 °C, against the standard’s ±1 °C. Three read higher inverted, at p < 0.0001. Those three swelled in water, and the authors judged the inverted method “less reliable for those dressings” for that reason. The other four showed no statistical difference, and the rankings didn’t agree across methods.
The widest gap was dressing F, at 2,838 upright and 30,530 inverted, about 10.8 times by our arithmetic. Its cup was found empty after the inverted test, and the authors call that value questionable. Every sample also turned concave over time, which they put down to vacuum in the cup.
| Method | EN 13726:2023 annex | What touches the dressing | What it simulates | Typical use |
|---|---|---|---|---|
| Upright | H, normative | Water vapor, across an air gap | A dressing over a lightly exuding wound or intact skin | Film and fixation dressings, including IV dressings |
| Inverted | I, normative | Liquid test fluid | A dressing over a wet, heavily exuding wound | Film dressings in the 2023 edition; foam dressings in studies using the 2002 method |
| Fluid handling capacity | E, normative | Liquid test fluid, for 24 hours | Total fluid managed: absorbed plus lost as vapor | Absorbent, foam and hydrocolloid dressings with a liquid-impermeable backing |
Table 1 of the standard lists film dressings against both MVTR annexes. Zehrer’s group ran foams inverted and upright under the 2002 method, and called the inverted value an estimate of “the theoretical MVTR of the dressing when placed over a highly exudating wound.” Their foams spanned 830 to 12,750 g/m²/24 h inverted against 80 to 1,620 upright. For IV films, Bainbridge’s group wrote: “We suggest that the upright method is the correct one to use for IV dressings.” They expect fluid pooling under such a dressing only in extremes “such as diaphoresis or high ambient temperature and humidity.”
How does MVTR relate to fluid handling capacity?
MVTR and fluid handling capacity are separate tests. EN 13726:2023 defines fluid handling capacity as fluid absorbed plus moisture vapor loss, the fluid transpired through the dressing during that test. Moisture vapor loss is the vapor half of the capacity, as an amount. MVTR is a rate, measured in its own annexes on films.
A film’s MVTR describes the film. A bordered foam’s fluid handling capacity describes the whole dressing, absorbent and backing together. Wound dressing fluid handling covers which dressings take the fluid handling test, and free swell versus absorption under load covers the absorption tests.
Where does ASTM E96 fit?
ASTM E96 is a general water vapor transmission method, written for materials rather than wound dressings. Its current scope covers “paper, plastic films, other sheet materials, coatings, foams” among others, and provides “Two basic methods, the Desiccant Method and the Water Method.” The user sets and reports the conditions. Its section 5.2.1 warns against using a result from one method or set of conditions to predict another.
Xu and colleagues cited ASTM E96 for their polyurethane membranes, run on a cup of distilled water at 37 °C and 50% relative humidity. Zehrer’s EN 13726 setup ran at 19%. Humidity sets the vapor gradient across the film, so a number from one setup doesn’t compare with a number from the other.
Which conditions does an MVTR test leave out?
An MVTR test leaves out the patient: movement, skin, real exudate and a whole dressing. Bainbridge’s group writes that the tests “do not provide the true MVTR experienced when a dressing is applied on skin.” Zehrer’s group compared seven foam dressings in the cup and on an artificial wound on volunteers’ backs, and found the two “only moderately correspond.” They list the limits themselves:
- Sample: a small cut piece, rigidly mounted, rather than a full dressing.
- Movement: the static test leaves out patient movement and the adhesive’s seal to skin.
- Conditions: 37 °C and 19% humidity, “conditions that probably do not reflect actual clinical conditions.”
- Fluid: water, rather than wound exudate.
- Volunteer model: it “has not been validated for replicating clinical levels of wound exudate.”
No target value settles the question in the sources checked here. Xu found a membrane near 2,028 g/m²/24 h worked best in their cell model and mouse wounds, and they note an earlier recommended range was “suggested indirectly” with “no detailed data from animal experiments.” The right MVTR depends on the wound, and your specification sets the limit.
What a GelTech study measures on MVTR
A GelTech dressing study runs MVTR upright and inverted on a test menu aligned with EN 13726. Carousel runs unattended and weighs each sample to 1 mg, six samples at a time. Clause 5 of EN 13726:2023 says any deviation from an annex’s test parameters is validated by the lab and noted in the report, so your report states the setup and conditions behind each number.
GelTech doesn’t run clinical or animal studies, or sterility or biocompatibility testing. See wound dressing testing, how a pilot study works and what to put in a study brief.
Sources
- CEN. EN 13726:2023 Test methods for wound dressings: aspects of absorption, moisture vapour transmission, waterproofness and extensibility. Scope, definitions 3.5 and 3.14 to 3.16, clauses 4 and 5, Table 1 and Annex A items 1 to 14, in the SIST EN 13726:2024 preview
- Bainbridge P, Browning P, Bernatchez SF, Blaser C, Hitschmann G. Comparing test methods for moisture-vapor transmission rate (MVTR) for vascular access transparent semipermeable dressings. J Vasc Access. 2023;24(5):1000-1007 (funded by 3M)
- Zehrer CL, Holm D, Solfest SE, Walters SA. A comparison of the in vitro moisture vapour transmission rate and in vivo fluid-handling capacity of six adhesive foam dressings to a newly reformulated adhesive foam dressing. Int Wound J. 2014;11(6):681-690 (funded by 3M)
- ASTM E96/E96M-24a Standard test methods for gravimetric determination of water vapor transmission rate of materials. Scope and significance (public summary page)
- Xu R, Xia H, He W, et al. Controlled water vapor transmission rate promotes wound-healing via wound re-epithelialization and contraction enhancement. Sci Rep. 2016;6:24596
Related reading
GelTech services and industries
Questions
Is a higher MVTR better for a wound dressing?
A higher MVTR isn't better for every wound dressing. It suits a heavily exuding wound, where the dressing has to shed fluid, and works against a dry wound that needs moisture kept in. The target comes from the wound type and the product specification, since EN 13726 sets no MVTR pass mark.
Does a bordered dressing's MVTR describe the whole dressing?
A bordered dressing's MVTR describes the film that was mounted on the cup. Bainbridge's group tested a section of film and left out any bordered or reinforced area, so an adhesive border or a reinforced edge wasn't part of the number.
Can an MVTR from ASTM E96 be compared with one from EN 13726?
An ASTM E96 MVTR and an EN 13726 MVTR shouldn't be compared directly. E96 leaves temperature and humidity to the user, who reports them, and it covers materials in general rather than dressings, so two numbers from the two methods rarely share conditions. Compare dressings tested by the same method, under the same conditions.