Where manual swelling and degradation tests vary
Manual gravimetric readings vary most at the hand steps between the bath and the balance: the blot, time in air, timing, transfers, drying and the person doing them. A contact lens standard, since withdrawn, named the blot as the main limit on its accuracy, and its ring test put a lens-sized sample at about ±1 percentage point of water content.
Where do manual swelling and degradation readings vary?
Handling varies at six steps, and a manual lab can tighten each one at some cost. The 1997 edition of the contact lens standard ISO 10339, since withdrawn, covers four of them in one procedure: the blot, time in air, drying and the operator.
Handling isn’t the only source of spread. Batch-to-batch differences in the material, the ions in the medium and the method itself have no handling fix. Zhang, Feng and Jin found the tea-bag method confines a swelling powder, “leading to the measured swelling capacity significantly lower than their actual capacity”.
| Step | What varies | Effect on the reading | How to reduce it |
|---|---|---|---|
| The blot | Pressure, paper or cloth, dampness, number of presses | Over-blotting reads low; a damp blot reads high | One written blot, and samples of 100 to 300 mg hydrated, which use more material |
| Time out of medium | Seconds in air between lift and weigh | A gel loses water and reads low; a swelling powder keeps taking up fluid and reads high | A timer, a fixed sequence, a closed or tared weighing vessel |
| Timing | Minutes between the planned and the actual reading | Wide scatter in the first minutes, where a slip is a larger share of elapsed time | Staggered starts, and plotting against the actual time |
| Transfers | Fragments and particles left behind or poured off | Mass reads low, and the loss stays out of every later reading | A holder or mesh that keeps the sample together, and catching what falls off |
| Drying | Residual water, cooling, a fixed clock | Depends on which dry mass holds the water (see below) | Drying to constant weight, which normally takes 16 to 18 hours in an oven, and cooling before weighing |
| Operator | Technique between people, and drift in one person | Offsets between operators, wider spread within one | Training against a reference sample, one operator per study |
How does the blot change a reading?
The blot can push a wet reading either way. ISO 10339 allowed a dry lint-free cloth pressed “three times with a finger tip” or “barely dampened” filter paper. It warned that the dry cloth risks over-blotting, “which could lead to an underestimate”, and that damp paper “may not remove all surface droplets” and could overestimate.
The standard’s ring test found three determinations on thick polymer discs within ±0.4% water content, and expected about ±1.0% for a lens-sized sample. By our arithmetic, that ±1 point matters more on a wetter gel: at 92% water a swelling ratio, (Ws − Wd)/Wd, is about 11.5 g/g, and 91% or 93% gives about 10.1 or 13.3 g/g.
Write the blot down: the material, the number of presses, and how hard. Keep the sample mass up where you can, since the standard set 100 to 300 mg of hydrated sample “to minimize the error in blotting technique”. A larger sample uses more material per timepoint. The formulas are on the hydrogel swelling ratio page.
How much does time out of the medium matter?
Seconds matter, and they can move the reading in either direction. ISO 10339 said “evaporation can also occur if the procedure is not carried out rapidly”, and asked for the blot and the weighing “as quickly as possible”. For superabsorbent powders, Zhang and colleagues timed fluid removal at 16 to 22 seconds with a tea bag and 5 to 7 seconds with their sieve filtration method. They note the powder can “continue absorbing liquid during the time of excessive fluid removal”.
A timer fixes the time out of the medium. Weigh the samples in the same order at every timepoint. A covered balance and a tared, closed vessel cut evaporation.
Why do early readings scatter?
Readings in the first minutes after immersion can scatter widely. In a round robin of 13 research groups testing two superabsorbent polymers for concrete, the abstract reports “high scatter” in the early minutes and gives no cause. Two are plausible: a one-minute slip is a larger share of elapsed time at five minutes than at five hours, and a powder keeps absorbing during fluid removal, as Zhang and colleagues note.
Stagger the start of each sample so its clock runs from its own immersion, record the actual time of every reading, and plot against that time rather than the planned one.
How are samples lost between readings?
Material a wet sample sheds in handling is gone from every later reading of that sample. Fisher and colleagues blotted in situ forming PLGA scaffolds containing hydroxyapatite before each weighing. They put the lower late wet mass down to small amounts of hydroxyapatite that were likely resuspended and removed with the supernatant at each timepoint. Polymer degradation testing in depots walks through that study.
A mesh basket or holder that travels with the sample keeps it together. Catch what falls off, and record the timepoint when shedding starts.
Why does drying to constant weight vary?
A dry mass varies when drying stops on a clock instead of at constant weight. ISO 10339 dried in an oven at 105 °C “until constant mass is achieved”, noted this “normally takes 16 h to 18 h”, and cooled the sample in a desiccator for at least 30 minutes before reweighing.
Which way the error runs depends on which dry mass holds the water. Water left in a final dry mass makes it read high, so mass loss reads low. Water left in the starting dry mass makes a swelling ratio read low and a mass loss read high. Degradation samples are often freeze-dried instead, which mass loss, molecular weight and erosion covers.
Weigh repeatedly until two readings agree, cool before every weighing, and dry and weigh in the same tared container.
How much do operators differ?
Operators differ between each other and from day to day. Holland and Davies describe variation from “the same person unknowingly performing a task differently each time”, as well as between individuals. ISO 10339 told the operator to “become proficient at the blotting technique before proceeding”. Zhang and colleagues built their error bars from at least nine measurements by three experimenters, so the spread between people sits inside them.
Train each operator against a reference sample until their readings match, keep one operator through a study, and record who ran each timepoint.
What does consistent handling buy?
Consistent handling narrows the spread that handling adds, which lets a smaller difference between formulations show. A ranking of two candidates holds only when both got the same blot, the same timing and the same drying. Comparing hydrogel formulations covers a fair comparison.
Automation repeats the handling, and it brings limits of its own. Holland and Davies warn of the “rapid propagation of errors” in automated tasks, that “an inherent machine to machine variability can remain”, and that it “can be more hidden” than in manual procedures. A drained wet mass and a blotted one aren’t directly comparable either, so a lab moving between them should run both on the same material before setting new numbers beside its history. What to automate first covers when to take the hand steps out.
What a GelTech study measures on handling variability
In a GelTech study, Carousel takes the repeated reading, so every timepoint gets the same handling. It tracks drained wet mass, weighed to 1 mg at every timepoint, for six samples at a time, each in a 24 mL sample chamber inside a 50 mL incubation chamber, and its mass data carries roughly 3x less data variability than manual work. Release and chemistry come from integrated instruments such as UV-Vis, HPLC and LC-MS.
The automated method is on automated hydrogel swelling measurement, and the tests GelTech runs are on fluid handling and degradation testing. GelTech doesn’t run animal or clinical studies, or accredited biocompatibility or sterility testing. See how a pilot study works.
Sources
- ISO 10339:1997 (withdrawn 2006, replaced by ISO 18369-4). Ophthalmic optics, contact lenses, determination of water content of hydrogel lenses (public preview, Annex A)
- Zhang K, Feng W, Jin C. Protocol efficiently measuring the swelling rate of hydrogels. MethodsX. 2020;7:100779
- Mechtcherine V, et al. Testing superabsorbent polymer (SAP) sorption properties prior to implementation in concrete: results of a RILEM round-robin test. Mater Struct. 2018;51:28 (abstract)
- Fisher PD, Palomino P, Milbrandt TA, Hilt JZ, Puleo DA. Improved small molecule drug release from in situ forming PLGA scaffolds incorporating poly(beta-amino ester) and hydroxyapatite microparticles. J Biomater Sci Polym Ed. 2014;25(11):1174-1193
- Holland I, Davies JA. Automation in the life science research laboratory. Front Bioeng Biotechnol. 2020;8:571777
Related reading
GelTech services and industries
Questions
Should a lab blot or drain a swollen gel before weighing?
A blot and a drain both remove surface fluid, and either works if it's done the same way every time. Draining through a fixed mesh for a fixed time takes less judgment than blotting, while blotting suits a single intact piece that holds its shape. Pick one and keep it for the whole study.
How many replicates does a manual swelling test need?
A manual swelling test needs enough replicates to see the spread that handling adds, and the right number depends on how big a difference you need to detect. Run a pilot with your own operator and material, and add samples if the spread between replicates is close to the difference you're trying to detect.
Can two labs' manual swelling numbers be compared?
Two labs' manual swelling numbers can be compared only when the formula, medium, temperature, timing and surface-fluid method all match. In a 13-group round robin on superabsorbent polymers for concrete, the tea-bag and filtration methods corresponded in behavior, filtration varied less in absorption capacity after 24 hours, and readings in the first minutes scattered widely.