What makes a biomaterials dataset useful

A biomaterials dataset is useful when someone else can rebuild every number in it: the material and batch, the conditions, the handling, the instrument, the timepoint, the units and the reference state, with raw readings kept beside processed values. FDA’s 2018 guidance for drug manufacturing names the checklist ALCOA: attributable, legible, contemporaneous, original or a true copy, and accurate.

Which ALCOA+ properties should a lab record have?

ALCOA is a checklist for whether a record can be trusted, and it was written for regulated GxP records. FDA’s 2018 guidance on data integrity for drug CGMP defines data integrity as the completeness, consistency and accuracy of data, which should be attributable, legible, contemporaneously recorded, original or a true copy, and accurate. It covers CGMP for drugs under 21 CFR parts 210, 211 and 212.

MHRA’s 2018 GXP guidance uses ALCOA too. It notes that a “+” was added later to emphasize four more properties, complete, consistent, enduring and available, and that expectations are the same whichever acronym you use. Two limits sit in its scope section. It doesn’t extend to medical devices, a class that takes in many implants and dressings, and it addresses integrity rather than quality, because integrity controls don’t necessarily guarantee the quality of the data generated.

Neither guidance was written for device development data. For an R&D team, that makes ALCOA+ a useful checklist rather than a compliance claim.

Property What it means A biomaterials example
Attributable Who or what made the record, and when Each swelling weight carries the operator or instrument ID and a timestamp
Legible Readable for the whole retention period A handwritten 0.52 g that can’t be confused with 0.32 g
Contemporaneous Recorded when the work happened The day-14 mass is logged on day 14, never reconstructed from memory on day 15
Original The first capture, or a true copy of it The balance’s own record, rather than the value retyped into a spreadsheet
Accurate Correct, with corrections traceable A mistyped batch number is corrected with the old value still visible
Complete Nothing dropped, failures included A disc that broke on day 21 stays in the record with a note
Consistent Same units, formats and sequence throughout Mass in mg in every file, and day 0 defined the same way in every study
Enduring Stored on media that last Raw data kept beyond the end of the project rather than on one laptop
Available Retrievable for review when needed A reviewer can pull the day-28 images for any chamber

What do the FAIR principles add?

FAIR asks whether data can be found and reused, by people and by software. Wilkinson and colleagues’ 2016 principles cover four properties, findable, accessible, interoperable and reusable, and put the emphasis on letting machines find and use data automatically.

Three of the sub-principles matter most for a lab:

  • Rich metadata: data described with many accurate, relevant attributes.
  • Provenance: where each value came from and what was done to it.
  • Community standards: shared terms and formats, so two labs’ datasets line up.

ALCOA asks whether a record is trustworthy. FAIR asks whether it’s reusable. A model, such as one in a self-driving lab, needs both.

Which metadata does a biomaterials dataset need?

Every reading needs enough context to be rebuilt. In FDA’s guidance, the number 23 means nothing without the unit mg, and the guidance lists timestamps, user IDs, instrument IDs and material identification numbers as metadata.

For a biomaterials study, record at least:

  • Material: composition, polymer grade and molecular weight, crosslinker and its concentration.
  • Batch: batch or lot number, preparation date and who made it.
  • Conditions: medium, pH, temperature, volume, agitation and any enzyme or surfactant.
  • Handling: how the sample was drained, blotted, dried or cut, and how long each step took.
  • Instrument: ID, resolution and calibration date.
  • Timepoints: elapsed time from a defined day 0.
  • Units: for every column.
  • Reference state: dry mass, initial wet mass or as-made mass, named in the file.

Processing history belongs here too. Lu, Garcia and Mikos found that PLGA films degraded during sample processing, so a film’s starting state depends on how it was made.

Why keep raw data beside processed data?

Processed values carry choices that only raw data can check. MHRA defines raw data as the first capture of information and says it must permit full reconstruction of the activities. Information captured in a dynamic state should stay available in that state, though a balance or pH meter that only prints makes its printout the raw data. MHRA also asks that the amount and resolution of data collected be justified. FDA gives an example: an FTIR spectral file can be reprocessed, while a printout of it can’t.

A swelling ratio comes from two masses and a formula, and labs use more than one formula. Keep every weighing, and the ratio can be recalculated another way. Keep only the ratio, and it can’t.

How should replicates be labelled?

Label each sample as a technical replicate or an independent sample. Vaux, Fidler and Cumming explain that replicates, measurements of one preparation, give n = 1 however many you make. They serve as internal quality checks, and independent samples, repeated as separate experiments, are what support a conclusion.

For a hydrogel, three discs cut from one batch tell you about measurement and handling spread. Three separately made batches tell you whether the formulation behaves the same way each time. A dataset that doesn’t say which it holds can’t support a claim about the formulation.

FDA’s guidance adds that in CGMP records, all data, including failing and suspect results, stays in the record, and invalidating a result needs a documented, scientifically sound justification. An R&D dataset that records excluded samples and the reason is easier to trust.

Why does consistent handling matter for comparing studies?

Two studies compare only if they handled samples the same way, because water held by the sample moves the result. Deng and Uhrich found that absorbed fluid changed PLGA’s storage modulus, and fluid uptake grew over six weeks. By our reasoning, the same holds for wet mass: the fluid left on the sample is part of the number, so drain time and blot pressure move it.

A fixed drain step has its own trade-off. It leaves a surface film, and the film can change as an eroding sample gains surface area and pores, as mass loss, molecular weight and erosion explains.

That’s why a fair comparison of hydrogel formulations holds handling fixed, and why a quality reviewer choosing a testing partner asks how each reading is taken.

What a GelTech study measures on dataset quality

A GelTech study handles every mass reading the same way. Carousel runs six samples at a time and records drained wet mass at every timepoint at 1 mg resolution, with pH and temperature monitored in line and an image of every sample. Its data comes out as .csv, and its images as .jpg. On mass data, GelTech reports roughly 3x less data variability than manual work. Independent batches each take a chamber, so a design with several batches per formulation fits fewer formulations into one run of six. Integrated instruments such as UV-Vis, HPLC, LC-MS, FTIR and rheometry read release, chemistry and mechanics. GelTech doesn’t run animal or clinical studies, or accredited biocompatibility or sterility testing. See the platform, what to automate first and how a pilot study works.

Sources

Related reading

GelTech services and industries

Questions

Is ALCOA a legal requirement for R&D data?

ALCOA is written into regulator guidance for GxP records, such as manufacturing and quality control data for drugs. FDA's 2018 guidance calls itself recommendations unless specific regulatory or statutory requirements are cited, and the record rules it explains, such as 21 CFR 211.68, 211.180, 211.188 and 211.194, are binding for drug manufacturing. Early R&D data often sits outside those rules, though data that later supports a submission may be read against them.

What's the difference between ALCOA and ALCOA+?

ALCOA+ adds four properties to ALCOA: complete, consistent, enduring and available. MHRA's 2018 guidance keeps the shorter acronym, says the '+' was added later to emphasize the requirements, and expects the same either way.

Can processed data replace raw data in a report?

Processed data can sit in a report, but it can't replace the raw data behind it. A reader with only processed values can't check the choices made along the way, such as which swelling-ratio formula was used or which samples were left out.

Does FAIR data have to be public?

FAIR data doesn't have to be public. Wilkinson and colleagues' principle A1.2 lets the retrieval protocol include authentication and authorization where necessary.